Initial experience of a large, self-expanding, and fully recapturable transcatheter aortic valve: The UK & Ireland Implanters' registry.
Dowling C, Firoozi S, Doyle N, Blackman DJ, Malkin CJ, Cunnington MS, Saraf S, Buch MH, Levy R, Chowdhary S, Spence MS, Manoharan G, Owens CG, Brennan PF, Roberts D, More R, Wiper A, Abdelaziz HK, Mylotte D, Neylon A, Martin N, Mercanti F, Dorman S, Panoulas V, Dalby M, Kashyap MN, Kabir T, Kovac J, Kontoprias K, Malik IS, Ghada MW, Sen S, Ruparelia N, Demir OM, Frame A, Uren NG, Anderson R, Rajathurai T, Tapp L, Deegan L, Grech E, Hall I, Neville M, Rampat R, Hildick-Smith D, Mullen M, Kennon S, Chandrala P, Doshi S, Brecker SJ.
Dowling C, et al. Among authors: more r.
Catheter Cardiovasc Interv. 2019 Mar 1;93(4):751-757. doi: 10.1002/ccd.27934. Epub 2018 Nov 5.
Catheter Cardiovasc Interv. 2019.
PMID: 30394657
Mean gradient was 7.0 4.6 mmHg and effective orifice area 2.0 0.6 cm(2) . Paravalvular regurgitation was more than mild in 7.2%. A new permanent pacemaker was implanted in 15.7%. Early safety was demonstrated in 91.2%. ...CONCLUSIONS: Real-world experience of the 34 mm Evo …
Mean gradient was 7.0 4.6 mmHg and effective orifice area 2.0 0.6 cm(2) . Paravalvular regurgitation was more than mild in 7.2%. A ne …