Two-parametric prescan calibration of gradient-induced sampling errors for rosette MRI

Magn Reson Med. 2024 Oct 22. doi: 10.1002/mrm.30355. Online ahead of print.

Abstract

Purpose: The aim of this study was to develop a simple, robust, and easy-to-use calibration procedure for correcting misalignments in rosette MRI k-space sampling, with the objective of producing images with minimal artifacts.

Methods: Quick automatic calibration scans were proposed for the beginning of the measurement to collect information on the time course of the rosette acquisition trajectory. A two-parameter model was devised to match the measured time-varying readout gradient delays and approximate the actual rosette sampling trajectory. The proposed calibration approach was implemented, and performance assessment was conducted on both phantoms and human subjects.

Results: The fidelity of phantom and in vivo images exhibited significant improvement compared with uncorrected rosette data. The two-parameter calibration approach also demonstrated enhanced precision and reliability, as evidenced by quantitative T 2 * $$ {\mathrm{T}}_2^{\ast } $$ relaxometry analyses.

Conclusion: Adequate correction of data sampling is a crucial step in rosette MRI. The presented experimental results underscore the robustness, ease of implementation, and suitability for routine experimental use of the proposed two-parameter rosette trajectory calibration approach.

Keywords: gradient imperfections; k‐space misalignment; rosette trajectory; trajectory estimation.